Chang-Hoon
Choi1, Ioannis Lavdas1, James M. S. Hutchison1,
David J. Lurie1
1Aberdeen Biomedical Imaging Centre,
University of Aberdeen, Aberdeen, Scotland, United Kingdom
An
off-resonance RF pre-saturation pulse is typically employed for magnetisation
transfer contrast MRI. Measuring the magnetisation transfer (MT) effect as a
function of magnetic field (B0)
may provide valuable. In order to conduct field-dependent MT experiments, two
techniques are required. Firstly, B0
should be switched between levels by fast field-cycling during irradiation of
the saturation pulse, and secondly the resonance frequency of the resonator (f0) should also be shifted
simultaneously. Here, we constructed the frequency-switchable RF coil using
PIN diodes. (f0) is
actively switched between five different values, with excellent impendent
matching (about -40dB) and the Q-factor (approximately 80).