is typically difficult to differentiate NAAG directly from NAA of standard
MRS due to the low concentration of NAAG and overlap with other metabolites.
Five repetitive scan trials of five phantom cases in which all the phantoms
were scanned using CSI at one-day intervals to figure out the reproducibility
and the accuracy of the measurement. The phantom cases contained a range of
concentrations of Glu, NAAG, and constant concentrations of other ten
metabolites. It was found that as the concentration of NAAG becomes smaller
(especially below 1mmol/kg), the overestimation bias in measuring the NAAG
gets stronger.