Frederik
Testud1, Iulius Dragonu1, Jrgen Hennig1,
Maxim Zaitsev1
1Medical Physics, Department of Diagnostic
Radiology,
GE
EPI is a widely used imaging technique, but is very sensitive to B0
field inhomogeneities. To correct for temporal changes of B0,
real-time measuring methods are necessary, such as estimating gradient maps
of B0 from raw EPI data. Here a filter scheme is presented to
calculate local B0 gradients. The local gradient in the readout
direction is estimated independently from the gradient in the phase encoding
direction by finding the contour lines of the gradients. This method is
compared with previously introduced raw data based techniques and shown to
perform better or equally well.