Hendrikus
Joseph Alphons Crooijmans1, Klaus Scheffler1, Oliver
Bieri1
1Division of Radiological Physics,
Department of Medical Radiology,
The
DESPOT2 theory is based on the assumption of instantaneous RF pulses.
However, this is a pure theoretical assumption and it can never be met in
practice, only approached with short pulse durations. Explicitly in cases
where MT effect reduction is desired, long RF pulses are applied and the
assumption is not met leading to deviation of calculated T2 from
true T2 values. The implementation of a correction for finite
pulse effects in the DESPOT2 theory makes the method independent of RF pulse
duration and marginal deviations of around 1% of the true T2 are
obtained for the calculated T2.