Feng Huang1,
Wei Lin1, Yu Li1, Arne Reykowski1
1Invivo Corporation,
It
has been shown that joint image reconstruction and sensitivity estimation in
SENSE (JSENSE) can improve image reconstruction quality when acceleration
factor is high. However, existing methods for JSENSE need long reconstruction
time and/or optimal termination condition, which have hindered its clinical
applicability. In this work, a fast non-iterative JSENSE technique, based on
pseudo full k-space, is proposed to improve the clinical applicability of
JSENSE. Using the proposed method, the computation time for sensitivity maps
could be reduced from minutes to a few seconds without degrading the image
quality.