low-flip-angle, phase-based method for calibrating the transmitter voltage
for hyperpolarized gas MRI has been presented in previous studies. This work
introduces two optimized versions of the phase-based method, and evaluates
their performance in the presence of B0 and B1 inhomogeneities compared to
that for an amplitude-based method. Results show that the accuracy of all
three methods is affected by significant B1 inhomogeneity; for significant B0
inhomogeneity, the amplitude-based method is robust while the phase-based
methods are very sensitive, particularly at relatively low flip angles.