Isotropic Resolution 3D Fast Spin Echo Acquisition for Quantitative Arterial Spin Labelled Perfusion Imaging in the Kidneys

Philip M. Robson1, Ananth J. Madhuranthakam2, David C. Alsop1

1Radiology, Beth Israel Deaconess Medical Center, Boston, MA, United States; 2Applied Science Laboratory, GE Healthcare, Boston, MA, United States


Most studies using arterial spin labelling (ASL) for perfusion in the abdomen have used 2D acquisitions in a limited number of slices. We evaluated 3D Fast Spin Echo (3D FSE) imaging for volumetric acquisition of perfusion in the kidneys. In sagittal image volumes over each kidney, isotropic 2.6-mm resolution was achieved allowing assessment in any orientation. Quantitative perfusion values were found to be comparable to a 2D ASL single-shot FSE sequence, and gave values for total renal blood flow that are in broad agreement with physiological values.