Derek K. Jones1
1CUBRIC,
Eddy currents plague diffusion MRI. When they produce a stretch
/ compression of the image along the phase encode direction, the resultant
change in voxel volume leads to a reduction/ increase in signal intensity.
Many eddy current correction packages fail to account for this signal change.
Here we show that the consequences can be drastic for diffusion tensor MRI,
with biases in fibre orientation being as big as 5 degrees in regions of low
anisotropy. We conclude that the signal intensity must be modulated by the
volumetric change, in order to obtain meaningful and robust results from
diffusion MRI.