A Composite Decoupling Method for RF Transceiver Array Coils in MRI

Yunsuo Duan1, Bradley S. Peterson1, Feng Liu2, Alayar Kangarlu3

1MRI Research, Department of Psychiatry, NYSPI & Columbia University, New York, NY 10032, United States; 2MRI Research, Department of Psychiatry,, NYSPI & Columbia University, New York, NY 10032, United States; 3MRI Research, Department of Psychiatry, NYSPI & Columbia University,, New York, NY 10032, United States


We present a composite scheme for minimizing the mutual inductance of transceiver array coils with more than two channels, which combines a novel shielding-based decoupling design with simplified capacitive decoupling method. The significant advantage of this composite strategy is that the adjustment of decoupling parameters for each coil element barely affects other elements. This characteristic of the new strategy greatly minimizes the coupling mechanism between different elements which is inherent to coil arrays and thereby simplifying the complex decoupling of transceiver array coils.