Correction of Eddy Currents for Time-Domain-Interleaved Blipped-Phase-Encoding Echo-Planar Spectroscopic Imaging

Yoshitaka Bito1, Koji Hirata1, Satoshi Hirata1, Toru Shirai1, Toshihiko Ebisu2, Yuko Kawai3, Yosuke Otake1, Yoshihisa Soutome1, Hisaaki Ochi1, Masahiro Umeda3, Toshihiro Higuchi4, Chuzo Tanaka4

1Central Research Laboratory, Hitachi, Ltd., Kokubunji-shi, Tokyo, Japan; 2Neurosurgery, Nantan General Hospital, Nantan-shi, Kyoto, Japan; 3Medical Informatics, Meiji University of Integrative Medicine, Nantan-shi, Kyoto, Japan; 4Neurosurgery, Meiji University of Integrative Medicine, Nantan-shi, Kyoto, Japan


High-speed spectroscopic imaging using the echo-planar technique is sometimes distorted by eddy currents. We developed an eddy current correction technique for time-domain-interleaved blipped-phase-encoding echo-planar spectroscopic imaging (TDI-BPE-EPSI). This technique uses correction of spatial shift due to chemical shifts in the blipped-phase-encoding direction before applying eddy current correction based on the water signal. Correction of eddy currents is demonstrated by applying this technique to a phantom and a rat brain in vivo. This technique is shown to be also useful in diffusion-weighted spectroscopic imaging, which causes more eddy currents due to strong diffusion gradients.