Rudolf Fritz
Fischer1, Kilian Weiss1, Christof Baltes1,
Markus Rudin1, Peter Boesiger1, Sebastian Kozerke1
1Swiss Federal Institute of Technology
and University
We
compare the sensitivity of the two spectroscopic imaging sequences Echo
Planar Spectroscopic Imaging (EPSI)
and Linear Response Equilibrium (LRE), a steady state free precession
sequence with intrinsic suppression of
periodic bands. Simulations and phantom experiments were performed
revealing a good SNR performance of LRE especially at low spectral bandwidth.