Meeting Banner
Abstract #0552

SNR Quantification with Phased-Array Coils and Parallel Imaging for 3D-FSE

Charles Qingchuan Li1, Weitian Chen2, Philip J. Beatty2, Anja C. Brau2, Brian A. Hargreaves1, Reed F. Busse3, Garry E. Gold1

1Radiology, Stanford University, Stanford, CA, United States; 2Global Applied Science Laboratory, GE Healthcare, Menlo Park, CA, United States; 3Global Applied Science Laboratory, GE Healthcare, Madison, WI, United States


Current clinical MRI techniques often employ parallel imaging, partial Fourier and multicoil acquisition to decrease scan time while maintaining image quality. To aid in image quality assessment, image noise statistics can be measured by reconstructing noise-only acquisitions through an identical linear pipeline as signal data, which may involve signal data-dependent steps such as parallel imaging, partial Fourier homodyne and multichannel reconstructions. In this study it was shown that SNR and CNR measurements performed in 146 clinical knee MRIs using this quantification method significantly differ from the measurements obtained using the traditional foreground and background volume of interest approach.