Christoph Barmet1, Bertram Jakob Wilm1,
Matteo Pavan1, Georgios Katsikatsos1, Jochen Keupp2,
Giel Mens3, Klaas Paul Pruessmann1
1Institute for Biomedical Engineering,
ETH and University,
A
higher-order concurrent field monitoring setup is introduced for routine head
MRI. It enables the tracking of dynamic field evolution up to 3rd order
concurrently with data acquisition. This is particularly important for
non-reproducible field contributions, e.g. due to magnet heating, breathing
or external fields. The field information allows for the correction of image
artifacts at the reconstruction stage.