Martijn Anton Cloos1, Nicolas Boulant1,
Michel Luong2, Guillaume Ferrand2, Christopher J.
Wiggins1, Eric Giacomini1, Alain France2,
Dennis Le Bihan1, Alexis Amadon1
1CEA, DSV, I2BM, NeuroSpin, LRMN,
Gif-sur-Yvette, France; 2CEA, DSM, IRFU, SACM, Gif-sur-Yvette,
France
Transmit-SENSE
gives the opportunity to implement short excitation pulses with good
flip-angle homogeneity at high field. For slice-selective pulses, this was
previously demonstrated using a spoke k-space trajectory. Here we present a
novel pulse design returning sub-millisecond pulses with excellent flip-angle
homogenization over an extended volume. Experimental results are shown at 7T,
demonstrating a 950-μs excitation pulse producing a 151.1 flip-angle
distribution over a 16-cm spherical phantom having the same electrical
properties as a human head.