Natalia
Gudino1, Matthew J. Riffe1, Lisa Bauer2,
Jeremiah A. Heilman3, Mark A. Griswold4
1Biomedical Engineering, Case Western
Reserve University, Cleveland, OH, United States; 2Physics
Department, Case Western Reserve University, Cleveland, OH, United States; 33Quality
Electrodynamics, Mayfield Village, OH, United States; 4Radiology
Department, Case Western Reserve University, Cleveland, OH, United States
We
present a Current-Mode Class-D (CMCD) Feedback amplifier with class-D
preamplification that avoids the characteristic DC losses of linear
preamplification. We demonstrated a good wave profile of the AM feedback
system that modulates the RF pulse and preliminary images that prove
successful operation of the system in the scanner.