Tae Hyung Kim1, Kee Chin Tan1, Song I. Chun1, Ki Sueng Choi1, Yong Hee Han1, Chi Woong Mun1
1Biomedical engneering, Inje Univ., Gimhae, Kyungnam, Korea
Hyperthemia therapy combined with Magnetic Resonance Imaging (MRI) has several
advantages for the treatment of small tumor. High Frequency Ultra Sound (HIFU)
method has been commonly used for heating source. HIFU method, however, has
some drawbacks. For this, coaxial-slot antenna driven by 2.45GHz -wave was
proposed. The coaxial-slot antenna was designed by computer simulation. To
optimize the design of antenna, reflection parameter(S11) and SAR distribution
in muscle model at the center frequency, 2.45, were analyzed using HFSS s/w.
The temperature was measured to test heating performances of hand-made
coaxial-slot antenna based on the distance from the heating point and power
using agarose gel phantom. The SAR distribution has similar trends on both
computer simulation and temperature measurement experiment. In this study
suggest that coaxial-slot antenna is a useful heating source for MR hyperthemia
therapy.