Meeting Banner
Abstract #4363

Improved Quantification of Pharmacokinetic Parameters at 3 Tesla Considering B1 Inhomogeneities

Robert Merwa1, Gernot Reishofer2, Thorsten Feiweier3, Karin Kapp4, Franz Ebner2, Rudolf Stollberger1

1Institute of Medical Engineering, Graz University of Technology, Graz, Austria; 2Department of Radiology, Medical University of Graz, Graz, Austria; 3MED MR PLM AW Neurology, Siemens AG Healthcare Sector, Erlangen, Germany; 4Department of Radiation Therapy, Medical University of Graz, Graz, Austria


The determination of kinetic parameters depends strongly on the inhomogenities of the RF-field. Due to the magnitude of these inhomogenities the values for the AIF and tissue concentrations are widespread which lead to an overestimation or underestimation of Ktrans and Ve. An essential improvement can be achieved if the dynamic data are corrected accordingly. The absolute difference of Ktrans and Ve obtained with the AIF in two comparable arteries can be improved by a factor up to 33 when using the correction procedure. Also the coefficient of variation of the kinetic parameters could be improved.