Meeting Banner
Abstract #4005

Computed Diffusion Weighted Imaging (CDWI) for Improving Imaging Contrast

Matthew Blackledge1, Ben Wilton1, Christina Messiou1, Dow-Mu Koh1, Matin O. Leach1, David J. Collins1

1CR UK Clinical Magnetic Resonance Research Group, Institute of Cancer Research and Royal Marsden Hospital, Sutton, Surrey, UK

Evidence is given that computed high b-value images based on ADC calculations present improved image quality over acquired images at the same high b-value. Two example cases are shown, the first demonstrating the improved SNR of computed over acquired b = 1400 s mm-2 images of iliac crest metastases and the second assessing the viability of extrapolating computed prostate images to b = 2000 s mm-2.