Meeting Banner
Abstract #3054

Evaluation of B1+ and E Field of RF Resonator with High Dielectric Insert

Chunsheng Wang1, Xiaoliang Zhang1,2

1Dept. of Radiology and Biomedical Imaging, University of California San Francisco, San Francisco, CA, USA; 22UCSF/UC Berkeley Joint Graduate Group in Bioengineering, San Francisco & Berkeley, CA, USA


<B>B1+</B> and <B>E</B> field inside a birdcage coil with and without a high-dielectric tube insert were investigated by finite difference time domain method. The high-dielectric insert can help to increase <B>B1+</B> field and decrease <B>E</B> field, which means high-dielectric insert has potential to increase signal to noise ratio.