Abstract #2152

"Between Weeks" Reproducibility of 3.0 Tesla Magnetic Resonance Spectroscopy for Measuring Hepatic Fat Content

Jochem Reinier van Werven1, J. M. Hoogduin2, A. J. Nederveen3, A. A. van Vliet4, P. Vandenberk5, E. S. Stroes6, J. Stoker3

1Radiology, Academic Medical Center, Amsterdam, Noord-Holland, Netherlands; 2Radiology, University Medical Center Utrecht, Netherlands; 3Radiology, Academic Medical Center Amsterdam, Netherlands; 4PBA International, Netherlands; 5Johnson & Johnson Medical BV, Belgium; 6Vascular Medicine, Academic Medical Center Amsterdam, Netherlands


Purpose: 1H-MRS is a non-invasive method to assess hepatic fat content (HFC), but there is sparse literature addressing between weeks reproducibility. Methods: In 24 subjects between weeks reproducibility was evaluated and assessed by Coefficient of Variation (CV), Repeatability Coefficient (RC) and Intraclass Correlation Coefficient (ICC)