Meeting Banner
Abstract #3553

A NOVEL QUANTITATIVE METHOD FOR MEASURING THE TORQUE ACTING ON CARDIOVASCULAR IMPLANTABLE DEVICES IN NMR SCANNERS

Canese R, D'Avenio G, Grigioni M, Podo F
Istituto Superiore di Sanit

A new method for the quantitative evaluation of the torque by which a cardiovascular device tends to be aligned along the static magnetic field is here proposed and validated. The method is based on a torsion balance, which enables accurate measurements of small forces. In our experimental set-up the method allowed measurement of torques as small as 10-5 N*m and is therefore applicable to small devices such as stents.