Tijl van der Velden1, Quincy van Houtum1, Mark W.J.M. Gosselink1, Peter R Luijten1, Vincent O Boer2, and Dennis W.J. Klomp1
1Radiology, UMC Utrecht, Utrecht, Netherlands, 2Danish Research Centre for Magnetic Resonance, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark
In this work a gradient
insert coil for breast MRI has been constructed. Its behaviour inside a 7T whole
body MR system was characterized using magnetic field cameras. Furthermore, the
possibility to correct eddy currents from the gradient insert coil using the
built-in gradient set has been investigated.