Reduced Distortion Artifact Whole Brain CBF Mapping using Blip-Reversed Non-Segmented 3D Echo Planar Imaging
Neville D Gai1, Yi Yu Chou1,2, Dzung Pham1,2, and John A Butman1
1Radiology & Imaging Sciences, NIH, Bethesda, MD, United States, 2Center for Neuroscience and Regenerative Medicine, Henry Jackson Foundation, Bethesda, MD, United States
Arterial
spin labeling is typically performed with segmented k-space acquisition schemes
to reduce B0 inhomogeneity related distortion. Non-segmented techniques offer the
advantage of higher SNR/time allowing greater
brain coverage in shorter scan times. Here we use a modified 3D EPI acquisition
scheme along with pseudo-continuous arterial spin labeling to correct for B0
inhomogeneity related distortion. By employing phase encoding along opposite
directions in alternating control-label pairs and with subsequent
post-processing, we correct for the distortion. CBF images were compared with
GM masks obtained from relatively distortion free MPRAGE images to show
improved localization of the CBF maps.
This abstract and the presentation materials are available to members only;
a login is required.