Rene Bastkowski1, Kilian Weiss1,2, David Maintz1, and Daniel Giese1
A novel method based on
single-voxel-spectroscopy (PRESS) for the analysis and correction of
eddy-current induced artefacts in spiral phase-contrast MRI is presented. It is
demonstrated, that 0th and 1st order corrections result in residual background
offsets of less than 0.5cm/s, inherently correcting for geometrical
misalignments between flow acquisitions as well as 2nd order spatial phase
offsets. The method does not require special hardware and can be applied as a
pre-scan.