The Potential for Eddy Current Induced Peripheral Nerve Stimulation from an Active Implanted Device Canister
Xin Chen1, Jonathan Edmonson2, and Michael Steckner1
1Toshiba Medical Research Institute USA, Inc., Mayfield Village, OH, United States, 2Medtronic CRHF, Mounds View, MN, United States
We used numerical simulations to investigate the potential for increased PNS likelihood with implanted device. Modeling of a gradient coil loaded with a human subject with a metallic implanted canister showed that the electric field around the device can increase by up to 3 fold, suggesting increased PNS likelihood.