Michael Childers1, Roya Hashemi Rad1, Richard Williamson1, and Shiloh Sison1
This abstract presents magnetically induced force
measurements per ASTM F2052 of materials commonly used in implantable leads.
Implantable leads which are constructed solely from
tested materials which pass the magnetically induced force testing acceptance
criteria (i.e. gravity force),
may not require magnetically induced force testing per
ASTM F2052 for MR conditionality with 3 T MR scanners.