Many factors
contribute to the inaccuracy of MR-guided breast biopsies. Significantly, the
lack of real-time visualization of tool advancement towards the biopsy site
increases their duration and rate of false negatives. In this work, a novel
approach for instrument tracking, relying on the spatial variation of the
magnetic field, and using a set of 3 axis accelerometers/gyroscopes/magnetic
field sensors, is presented. One dimensional
tracking with 1.3mm rms error was demonstrated in the fringe field of a 3T
magnet.