Zhoujian Li1, Sajad Hosseinnezadian1, Geneviève Guillot1, Georges Willoquet1, Laurène Jourdain1, Marie Poirier-Quinot1, Luc Darrasse1, and Jean-christophe Ginefri1
We have implemented a piezo-motor based
automation system for contactless impedance matching of a monolithic Transmission
Line Resonator (TLR) operating at 4.7 T. The automation system successfully achieved inductive matching to more
than -30dB of the TLR inside the magnet and no artifacts was observed on the
image of a rectangular box-shaped water phantom. A second image, acquired
in the same condition but in the presence of another piezo-motor fixed on a
side of the sample revealed that the close proximity of the piezo-motor to the
sample brings B1-field inhomogeneity.