Stefan HG Rietsch1,2, Stephan Orzada1, and Harald H Quick1,2
1Erwin L. Hahn Institute for MR Imaging, University of Duisburg-Essen, Essen, Germany, 2High Field and Hybrid MR Imaging, University Hospital Essen, Essen, Germany
In order to allow for improved SNR and higher
acceleration during image acquisition in the body at 7T, we present a coil with
8Tx/Rx microstrip line elements with meanders and 24Rx loop elements. This coil
comprises 8 building blocks each consisting of one Tx/Rx element and 3
overlapping loops which are actively detuned during transmit. With about
-19 dB reflection and an average decoupling of more than -30 dB, the
SNR can be boosted by about 21% in the abdomen. Evaluation of g-factors as well
as in vivo images of healthy volunteers in both abdomen and heart show
promising results.