SERIAL Excitation with Parallel Reception Allows Human Brain Imaging at 9.4 Tesla at Low Power and with Acceptable Image Uniformity across the Full Field of View

Keith R. Thulborn1, Chao Ma2, Ian C. Atkinson1, Theodore C. Claiborne1, Steven M. Wright3, and Reiner Umathum4

1Center for Magnetic Resonance Research, University of Illinois, Chicago, IL, United States, 2Beckman Institute, University of Illinois at Urbana-Champaign, Urbana, IL, United States, 3Department of Biomedical Engineering, Texas A&M University, College Station, TX, United States, 4Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany

SERIAL excitation produces uniform image intensity at low power at ultra high field but has not been applied to humans. FLASH sequences modified for sequential single coil excitation while retaining full array receive mode were combined with generalized total variation regularized SENSE reconstruction and 4- and 8- arrayed coils. Images with acceptable uniformity, contrast and resolution over the in vivo human brain are demonstrated at 9.4T using low power.