Pavan poojar1, Bikkemane Jayadev Nutandev2, Nithin N Vajuvalli1, C.K. Dharmendra Kuman2, Ramesh Venkatesan3, and Sairam Geethanath1
1Medical Imaging Research Centre, Dayananda Sagar College of Engineering, Bangalore, India, 2Bangalore, India, 3Wipro-GE Healthcare, Bangalore, India
In dynamic scans, the significant values of k-space dependent on the shape of the organ which leads to arbitrary k-space trajectories. Gradient optimization for arbitrary k-space trajectory using active contour is a new acquisition technique that has been applied on six DCE breast data. The arbitrary k-space trajectory was obtained by active contour and gradients are optimized by employing convex optimization based on hardware constraints. Image reconstruction was performed using Fourier transform with density compensation. $$$K^{trans}$$$ and Ve maps were generated for different acceleration factors (1x, 2x, 3x, 4x and 10x) on tumor region to demonstrate utility of the method.