Yi Wang1, Xingfeng Shao1, Thomas Martin1, Steen Moeller2, Essa Yacoub2, and Danny JJ Wang1
1Neurology, UCLA, Los Angeles, CA, United States, 2Center of Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States
The
application of balanced SSFP (bSSFP) is limited by the banding artifact
resulting from its sensitivity to field inhomogeneity. A common approach
for band reduction involves multiple measurements with different RF phase
cycling, at the cost of lengthened total imaging time. In this work, we present
a novel time-efficient bSSFP banding reduction technique by utilizing
simultaneous multi-slice (SMS) imaging with CAIPIRINHA to acquire multiple
phase-cycled images within the same imaging time of a single-band bSSFP scan.
Effective band reduction is demonstrated in phantom, abdominal and brain
imaging with SMS factor up to four.