Xiaobing Fan1, Shiyang Wang1, Milica Medved 1, Tatjana Antic2, Serkan Guneyli 1, Aytekin Oto 1, and Gregory S Karczmar 1
1Radiology, University of Chicago, Chicago, IL, United States, 2Pathology, University of Chicago, Chicago, IL, United States
Accurate
measurements of the arterial input function (AIF) are needed in pharmacokinetic
models to analyze dynamic contrast enhanced (DCE) MRI data. The AIF often cannot
be accurately measured due to T2* and water exchange effects. Therefore, population
AIFs are often employed in pharmacokinetic modeling. Here we report a new
8-parameter empirical mathematical model (EMM) that fits the AIF measured
directly from the external femoral artery after a dose of contrast agent that
was greatly reduced to minimize artifacts. The results showed that the EMM-AIF accurately
models both 1st and 2nd passes of contrast agent circulations.