A new, highly effective slicewise motion correction for BOLD MRI: SLOMOCO
Erik Beall 1 and Mark Lowe 1
1
Imaging Institute, Cleveland Clinic,
Cleveland, OH, United States
Head motion artifact is a major unsolved problem for
fMRI and fcMRI. The primary reason why current methods
do not solve motion is the limitation of
volume-synchronized or -interpolated motion. Real motion
happens at the slice level, and this has been shown to
seriously limit accuracy of volumetric measures.
Improved correction requires knowledge of slice-level
motion, and in a major advance for the field, we present
an algorithm that obtains this. Using a motion-injection
sequence to acquire realistic motion-corrupted data in
cadavers, we show the algorithm obtains the known motion
parameters with accuracy on the level of external
trackers.