Tissue-specific Variable Flip Angle of 3D Turbo-Spin Echo on 3.0T system: Optimized with Individual T1 and T2 value for Prostate.

Katsusuke Kyotani 1 , Yoshiko Ueno 2 , Satoru Takahashi 2 , Yu Ueda 1 , Tomoyuki Okuaki 3 , Nobukazu Aoyama 1 , Kazuhiro Kitajima 2 , Hideaki Kawamitsu 1 , and Kazuro Sugimura 2

1 Division of Radiology, Kobe University Hospital, Kobe, Hyogo, Japan, 2 Department of Radiology, Kobe University Graduate School of Medicine, Kobe, Hyogo, Japan, 3 Philips healthcare Asia pacific, Minato-ku, Tokyo, Japan

Our aim was to evaluate the usefulness of tissue-specific VRFA 3D-TSE optimized with measured T1 and T2 values of prostate gland comparing to standard 3D-TSE for T2W imaging. Contrast ratio between PZ and PZ cancer of the tissue-specific VRFA 3D T2W-TSE was significantly higher than that of standard 3D T2W-TSE (P < 0.05). Contrast ratio between PZ and TZ of the tissue-specific VRFA 3D T2W-TSE was significantly higher than that of standard 3D T2W-TSE (P < 0.05). Spatial resolution ratio of tissue-specific VRFA 3D-TSE was improved from standard 3D-TSE (P < 0.05)