Meeting Banner
Abstract #2930

In-vivo 37 Cl Magnetic Resonance Imaging at 7 Tesla

Anna Kollefrath 1 , Manuela Rsler 1 , Reiner Umathum 1 , and Armin M. Nagel 1

1 Medical Physics in Radiology, German Cancer Research Center (DKFZ), Heidelberg, Germany

With a customer made solenoid coil 37 Cl imaging at a 7T whole- body MRI system at a frequency of 24.3 MHz was performed although the natural abundance of 37 Cl is lower than that of the for measurements preferred 35 Cl. For phantoms with 0.9% sodium chloride solution and different agar concentrations the relaxation times (T2 * and T1) were determined and compared to the relaxation times of 35 Cl. The relaxation times are longer than the ones of 35 Cl. In-vivo imaging of a narcotized rat showed the feasibility of MRI using the 37 Cl isotope in comparison with 35 Cl and showed high signal intensities.

This abstract and the presentation materials are available to members only; a login is required.