Abstract #2325
Evaluation of an novel RFID-based navigation system for MRI-guided interventions and surgery
Felix V. Gttler 1 , Andreas Heinrich 1 , Michael Sonnabend 1 , and Ulf Teichgrber 1
1
Department of Radiology, University Hospital
Jena, Jena, Thuringia, Germany
The purpose of the study is to evaluate the suitability
of a novel RFID-based tracking system for intraoperative
MRI. Therefore the spatial accuracy and SNR according to
the NEMA standard MS 1-2008 was quantified. The
installation of an RFID system including transponders
and receivers in the magnet room in close distance to
the magnet has low of non-relevant influence on MRI.
However the spatial accuracy have to be improved for an
application as tracking system in intraoperative MRI.
This abstract and the presentation materials are available to members only;
a login is required.
Join Here